Fume hood variable air volume control system adopts a controller that integrates a fast actuator and a
controller. It monitors the opening of the fume hood window through displacement sensors and quickly
adjusts the opening of the air valve to achieve a constant air velocity on the control surface. It is suitable
for fume hood systems using butterfly valves or Venturi valves. When the exhaust volume is set below
the safe value, the system will activate the built-in sound and light alarm to provide maximum safety
protection for laboratory staff. Suitable for various types of experimental fume hoods and exhaust equipment.
There are multiple working mode settings that can meet the different needs of customers.
System accessory description:1. variable air volume valve:
Integrated design of controller and actuator, simple structure;2S fast actuator, Venturi uses 3S straight
stroke actuator, fast response speed(Response within 1S, stable 2~3S);The Venturi nozzle shape is used
to measure the wind structure, the air volume is accurate and stable, and the error is less than plus or
minus 4% air volume adjustment ratio 16:1 (there is a third-party test report);
2.controller
Controller and actuator integrated design, simple structure, convenient installation,The controller provides:
lighting control, status output, 485 communication (BMS communication port, support Modbus RTU communication
protocol) and other functions.
3.Displacement sensor
Measuring range: 1000MM----5KΩ Measuring accuracy: plus or minus 0.1%, connection: two-wire system,
not positive or negative. 0.6 Stainless steel plastic-coated wire rope;Select different models according to the
site conditions.
4.Control panel
(1) Power supply: DC 5V, ± 10%,<2W
(2) Working environment: 0...+50 ºC, 10... 95% RH (no condensation)
(3) Storage environment: -20...+50 ºC, 10... 95% RH (no condensation)
(4) LCD screen: IPS LCD screen, resolution ratio:1280 x 720
Fume Hood Ventilation System
We offer a variety of top quality specialized accessories of lab furniture and
fume hood including items below:
1. Various size plastic covers for steel frame.
2. Adjustable feet for lab cabinets.
2. Electric sockets with or without standout used on lab bench.
3. aluminum long trunking placed on worktop of lab bench.
4. Steel/stainless steel/PP hinge for wood/steel/PP cabinet respectively.
5. 304 stainless steel/Aluminum/Plastic grab handle for drawers of lab cabinet.
6. 3-joints drawer slide without noise.
6. LCD controller for fume hood in English.
7. Molded exhaust hood for fume hood.
8. Auto sash system for fume hood.
9. Whole VAV system for fume hood.
Trade information:
1. Delivery time: 6-20 days after getting advance payment.
2. Trade terms: EXW, FOB, C&F, CIF.
3. Payment terms: T/T, L/C, cash etc.
4. Port: Shanghai or any specified port in CHINA.
5. Packing: carton or wooden case.
6. Warranty period: 2 years once receiving our goods. Free accessories will be offered in case quality
problem happens within 2 years.
7. Service: online service to offer technical support or deal with any claims within 1 hour.